Fluke – 5502E Multi-Product Calibrator
מאפיינים מרכזיים
The 5502E has a wide range of outputs available from a single calibration source. These include dc and ac voltages up to 1020 V, dc and ac current up to 20.5 A directly or 1025 A (using the 5500A/COIL 50-turn current coil accessory) and variable resistances up to 1100 MΩ.
Using the 52120A Transconductance amplifier, the 5502E's output current is extended from 20.5 to 120 A; and with the use of 25 and 50 turn coils, it can calibrate instruments requiring currents up to 6000 A.
Choose the Fluke Calibration 5502E for easy, economical calibration of:
- Handheld digital multimeters (analog and digital) to 4.5 digits
- Bench multimeters (to 4.5 digits)
- Clamp meters
- Panel meters
- Electronic thermometers
- Chart recorders
- XY recorders
- Data loggers
Summary Specifications |
|
Function |
Range |
DC voltage | 0 to ± 1020 V |
DC current | 0 A to ± 20.5 A |
AC voltage | 1 mV to 1020 V 10 Hz to 500 kHz |
Volt*Hertz | 1000 V@ 10 kHz/33 V@100 kHz |
AC current | 29 μA to 20.5 A 10 Hz to 30 kHz |
Waveforms | Sine, square, triangle, truncated sine |
Resistance | 0 Ω to 1100 MΩ |
Capacitance | 220 pF to 110 mF |
Thermocouple (source and measure temperature) | B, C, E, J, K L N R, S, T, U 10 μV/°C and 1 mV/degC |
RTD (source temperature) | Pt 385-100 Ω, Pt 3926-100 Ω Pt 3916-100 Ω, Pt 385-200 Ω, Pt 385-500 Ω, Pt385 1000 Ω, PtNi 385-120 Ω, (Ni120), Cu 427 10 Ω |
Interfaces | RS-232, IEEE 488 |
Frequency | .01 Hz to 2 MHz |
לקבלת מידע צור קשר