Fluke – 52120A Transconductance Amplifier

מאפיינים מרכזיים

The Fluke Calibration 52120A Transconductance Amplifier supplies dc current up to 100 amps and ac current up to 120 amps, at accuracies to 140 ppm. Using accessory coils, it can generate 3,000 or 6,000 amps. Three 52120As connected in parallel can output up to 360 amps. Inductive drive capability of 1 mH and compliance voltage of 4.5 volts support a wide range of applications.

  • Industry-leading amplifier accuracy:
    • 140 ppm ac (used with 61XXA Electrical Power Standard or 5730A)
    • 350 ppm ac (used with dc/lf calibrator
    • 160 ppm dc (used with dc/lf calibrator
  • Frequency: DC to 10 kHz
  • Burden voltage (compliance): 4.5 V @ 120 A
  • Inductive drive capability: 1 mH load
  • Output ranges: 2 A, 20 A, 120 A
  • Input ranges: 2 V or 200 mA F.S. for 2 A and 20 A ranges, 1.2 V or 120 mA F.S. for 120 A range
  • Parallel operation: (Closed loop mode with 61XXX or 5730A) 2 or 3 amplifiers, up to 360 A in a single phase
  • Parallel operation: (Standalone mode), up to 10 amplifiers, up to 1200 A in a single phase
  • Accessory coils: 25-turn coil supports 3000 A, 50-turn coil supports 6000 A
  • Control communication with 6105A/6100B Electrical Power Standards or 5730A
  • GPIB remote operation
Operating limits
Output ranges 2 A, 20 A, 120 A (100 A dc)
Input current range 200 mA, 200 mA, 120 mA (100 mA dc)
Current gain 10, 100, 1000
Input voltage range 2 V, 2 V, 1.2 V (1.0 V dc)
Transconductance 1,10,100 Siemens
Frequency To 10 kHz
Maximum output compliance voltage 4.5 Vrms (6.4 Vpeak)
Inductive drive capability: 1 mH load
Output isolation, current terminal to earth 600 Vrms, 850 Vpeak, dc to 850 Hz
Performance specifications
AC accuracy, closed loop with 6105A To 140 ppm
DC accuracy, standalone To 160 ppm
AC accuracy, standalone To 350 ppm
Phase angle accuracy 0.006°
Load dependent phase angle shift <0.001° @ 60 Hz
General specifications
Input line voltage range 100 V to 240 V with up to ±10 % fluctuations
Transient overvoltage Impulse withstand (overvoltage); Category II of IEC 60364-4-443
Frequency 47 Hz to 63 Hz
Maximum consumption < 1500 VA
Calibration Documentation Report of calibration with data included; 17025 accredited report optional
Dimensions with feet (H x W x L) 192 mm x 432 mm x 645 mm
(7.6 inches x 17 inches x 25.5 inches)
Dimensions without feet (H x W x L) 178 mm x 432 mm x 645 mm
(7 inches x 17 inches x 25.5 inches)
Weight 25 kg (54 lb.)
Design standards and compliance Designed to EN 61010-1: 2010, CAN/CSA 22.2 No 61010.1-04, ANSI/UL 61010-1:2004, EN 61326-1:2006 CE marked, CSA listed
Operating temperature 5 °C to 35 °C
Calibration temperature range 16 °C to 30 °C
Warm up time 1 hour
Safe operating max. relative humidity (non-condensing) <80 % 5 °C to 31 °C ramping linearly down to 50  % at 35 °C
Operating altitude 0 m to 2,500 m
Shock and vibration MIL-PRF-28800F class 3
Specification confidence level 99%
לקבלת מידע צור קשר השאירו פרטיכם ונחזור אליכם